User Manual 

Sapphire Series® Static™ 

Low-Air-Loss Mattress Replacement System

Содержание Static Sapphire Pulse

Страница 1: ...User Manual 8083 Rev 8 0 03 04 2021 User Manual Sapphire Series Static Low Air Loss Mattress Replacement System...

Страница 2: ...rence 12 Patient Care Functions 13 Placing the Patient on the Mattress Surface 13 Positioning the Patient 13 Backrest Up or Fowler Position 13 Prone Position 14 Bedpan Placement Removal 14 Removing th...

Страница 3: ...2 Warranty Provisions 32 User Assistance Information 33 Warranty Information 32 Return Exchange Goods Policy 32...

Страница 4: ...he user to the presence of important operating maintenance or servicing instructions Disregarding a warning could result in patient and or user injury as well as damage to equipment Electrical Shock H...

Страница 5: ...perly protected from electrical shock This symbol marks the location of the leakage test point screw The hazards and warnings are indicated on the shipping container by this label This symbol marks th...

Страница 6: ...damage or failure Monitor the unit to ensure that it functions in the proper operating temperature CAUTION Keep out of direct sunlight CAUTION DO NOT store in temperatures below 0 F 18 C or above 95...

Страница 7: ...t the foot of the bed or sits on the floor under the bed The Sapphire Series Static Low Air Loss Mattress Replacement System has a detachable hospital grade electrical cord and a control panel with se...

Страница 8: ...d Frequency 220 Volts Input Power 50Hz 230 Volts Degree of Shock Protection Type B Maximum Relative Humidity 95 Storage Temperature 18 C to 35 C 0 F to 95 F Environmental Conditions Product must be st...

Страница 9: ...ght company and file a damage complaint immediately Environmental Conditions CAUTION Keep out of direct sunlight DO NOT expose to temperatures greater than 35 C 95 F or below 18 C 0 F DO NOT expose th...

Страница 10: ...attress with the Sapphire Series Static Low Air Loss Mattress Replacement System Secure air mattress to the bed frame with straps provided Warning or Safety Instructions Relating to Set Up WARNING 110...

Страница 11: ...d is unobstructed 6 NOTE Plug the 110V unit ONLY into grounded 110V AC outlet Plug the 220V unit ONLY into grounded 220V AC outlet 7 Turn master power switch ON located on the side of the unit 8 Press...

Страница 12: ...utes if there are no changes to the blower settings the lockout feature will activate b Press and hold the Lockout button for 3 seconds to disengage the lockout feature 5 Quick Deflation or CPR Use a...

Страница 13: ...3 sec Alarm Silence Mutes the audible alarm Visual light will not turn off until failure is resolved Autofirm Quickly inflates mattress to maximum firmness Fowler Used when head section of bed is elev...

Страница 14: ...art Association for performing CardioPulmonary Resuscitation CPR recommend a hard level surface for performing CPR moving the person to the floor if possible For performing CPR place the CPR board low...

Страница 15: ...ay be adjusted to increase the firmness of the center section after the pan is placed in position When the bedpan is to be removed logroll the patient off the bedpan and remove it Readjust the firmnes...

Страница 16: ...n the manufacturer s instructions All cleaning solutions must be diluted in accordance with manufacturer s instructions All covers must be rinsed after every cleaning cycle Rinsing of the support surf...

Страница 17: ...always be used as directed by the disinfectant s Safety Data Sheet 1 Prepare the disinfectant according to the manufacturer s recommendations 2 Prepare a separate bucket of warm fresh water to be used...

Страница 18: ...used for rinsing the equipment after cleaning disinfecting procedures are completed as instructed 3 Allow the air cells to remain wet with disinfectant solution for the manufacturer s recommended con...

Страница 19: ...ons to remove organic material and visible soil 6 Allow the control blower unit to remain wet with disinfectant solution for the manufacturer s recommended contact time 7 Rinse all surfaces of the con...

Страница 20: ...ain and service this product Inspect top cover for punctures rips tears or damage Inspect mattress base for punctures rips tears or damage Connect the control unit and verify proper operation if insta...

Страница 21: ...nction with the manufacturer supplied foam filter DO NOT attempt to install HEPA filter without the original equipment foam filter NOTE Ensure factory open cell foam filter is completely dry before pl...

Страница 22: ...oting Storage If you store your product for more than 30 days we recommend Remove HEPA filter Place HEPA filter in an air tight plastic bag or plastic wrap ensuring filter is sealed to prevent exposur...

Страница 23: ...flat on a clean level surface Avoid storage of other equipment on top of the support surface DO NOT expose the blower unit to humidity greater than 95 End of life Sunflower Medical L L C products must...

Страница 24: ...ury as well as damage to equipment or other property The use of cables or accessories other than those for which the blower was designed or tested can significantly degrade emissions and immunity perf...

Страница 25: ...gnetic radiation over the frequency range 80 MHz to 2 5 GHz The test was conducted on this blower and passed the requirements of IEC 61000 4 3 All pins of connectors have passed ESD testing List of Ca...

Страница 26: ...ent Guidance RF Emissions CISPR 11 Group 2 The blower must emit electromagnetic energy in order to perform its intended function Nearby electronic equipment may be affected RF Emissions CISPR 11 Class...

Страница 27: ...ality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 2kV Common A Mains power quality should be that of a typical commercial or hospital environmen...

Страница 28: ...equipment should be separated from the blower by no less than the distances calculated listed below D 3 5 V1 Sqrt P D 3 5 V1 Sqrt P 80 to 800 MHz Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz V...

Страница 29: ...olled The customer or user of the blower can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the blower as recom...

Страница 30: ...l entrapment as serious injury can result These guidelines can be found on the FDA website When using an Air Therapy system the care provider is responsible for ensuring the mattress properly fits the...

Страница 31: ...t punctured kinked or disconnected Ensure air cells are not punctured Check for proper connections from the hoses to the blower Make sure they are secure Check or clean air filters Ensure the CPR valv...

Страница 32: ...l units Part Number 27000445 Connector Connects from mattress to blower control unit side panel Allows air movement to inflate mattress Part Number 27400035 Part Number 27400049 Part Number 27400050 H...

Страница 33: ...Sunflower Medical LLC Warrants that all products provided under this agreement are free from defects in material and workmanship for the following stated warranty period from the date of delivery Sun...

Страница 34: ...ser Assistance Information For questions or assistance with this product contact Sunflower Medical L L C at Sunflower Medical L L C 206 Jefferson Street PO Box 276 Ellis Kansas 67637 Phone 1 888 321 3...

Страница 35: ...medical com Sunflower Medical L L C and Sunflower Medical L L C Rentals are trademarks of Sunflower Medical L L C Rentals L L C All specifications equipment and prices are subject to change without no...