28
Table 206 Recommended Separation Distances between portable and mobile RF
Communications equipment and blower.
Equipment and Systems that are NOT Life-Supporting
Recommended Separations Distances for the blower: The blower is intended for use in the
electromagnetic environment in which radiated disturbances are controlled. The customer or
user of the blower can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF Communications Equipment and the blower, as
recommended below, according to the maximum output power of the communications
equipment.
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P)
80 to 800 MHz
D=(7/E1)(Sqrt P)
800 MHz to 2.5 GHz
Compliance Level Cond RF 3 Rad RF-800 MHz 3 Rad RF – 2.5 GHz 3
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33