Changing the UV Lamp
The intensity of the UV lamp, particularly the emission of short-wavelength radiation, decreases over
time. Replace the lamp after every six months of operation. Instructions and illustrations for this
proced
ure also can be found in the Sensor’s
Quick Setup Guide
.
Warning
Installation of the UV lamp requires access to the inside of the
Sensor. To avoid potentially dangerous shock, disconnect the
power cord and remove the battery
before opening the Sensor’s
case.
Warning
The UV lamp contains mercury and may be considered hazardous material in
your local area. Dispose of the UV lamp in accordance with federal, state, or
local government regulations.
Warning
Should the UV lamp become broken or damaged, it should be handled in
accordance with your organization's toxic waste handling procedure and
disposed of in accordance with federal, state, or local government regulations.
N
OTE
: A cotton glove for wearing during the installation is included with
the replacement UV lamp. Wear this glove to avoid leaving fingerprints
on the quartz window of the lamp. Fingerprints absorb UV radiation and
decrease the performance of the oxidation reactor. If necessary, use
methanol to remove any fingerprints before installing the new lamp.