CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018
50 of 112
DLM 97200-03 EN Rev. A
7. Remove the covers from the two fuse holders and then remove the two fuses (see
Figure 7:
Preparing to Remove the Fuse Holders).
8. Remove the fuse holders by loosening the two screws and pulling the fuse holders out of the
Sensor.
9. Cut the tab connectors off the wires that connect to the right side of the fuse holders (see
Preparing to Remove the Fuse Holders).
Figure 7: Preparing to Remove the Fuse Holders
10. Strip the wires 1/4 in. (6.35 mm).
11. Secure the power terminal block to the Sensor by tightening the two screws.
12. Connect the existing wires to the power terminal. When viewed from the right side of the Sensor,
neutral (blue wire) connects to the left port and line (brown wire) connects to the right port (see
Attaching the Existing Power Wires). For each wire, insert the wire into the port and
then use a small screwdriver to tighten the screw on the top of the terminal block to secure the
wire.