CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018
76 of 112
DLM 97200-03 EN Rev. A
Storing the Sensor
When the Sensor will not be used for a prolonged time, the Sensor should be stored dry. Place the
Sensor in
Grab Sample
mode, make sure there is no sample water flowing to the Sensor, and begin a
Grab analysis to remove water from the Sensor. Alternatively, the
Pump
option on the
Diagnostics
screen can be used without adding any measurements to the data history.
Remove the battery from the Sensor.
Printing
The Sensor will automatically detect if a printer is connected and will automatically print results for
online, grab, cond cal, and TOC cal when they complete. When operating in the On-Line mode, the
printer prints out the variables shown on the On-Line display screen. For example, if you are looking
at an On-Line display screen that is showing only the TOC result, then the time and only the TOC
result prints.