DLM 97200-03 EN Rev. A
6
TOC calibration write
Problem encountered writing TOC calibration data to
flash. If the problem persists you may need to
change the system board. Contact Technical Support
or your local service provider.
7
TOC match write
Problem encountered writing TOC match data to
flash. If the problem persists you may need to
change the system board. Contact Technical Support
or your local service provider.
8
Auto-zero write
Problem encountered writing Auto-zero data to flash.
If the problem persists you may need to change the
system board. Contact Technical Support or your
local service provider.
9
Auto-offset write
Problem encountered writing Auto-offset data to
flash. If the problem persists you may need to
change the system board. Contact Technical Support
or your local service provider.
10
Flow rate calibration write
Problem encountered writing flow rate calibration
data to flash. If the problem persists you may need to
change the system board. Contact Technical Support
or your local service provider.
11
Data history queue overflow
Problem encountered writing data history to flash. If
the problem persists you may need to change the
system board. Contact Technical Support or your
local service provider.
12
Service data queue overflow
Problem encountered writing service data to flash. If
the problem persists you may need to change the
system board. Contact Technical Support or your
local service provider.
13
UV lamp should be replaced soon
Lamp expiration date is 10 days or less away.
14
Pump tubing should be replaced
soon
Pump tubing expiration date is 10 days or less away.
15
Battery needs recalibration
Battery is indicating that it needs to be recalibrated.
Put battery in external charger and press the
Recalibrate
button.
16
Invalid lamp install date
Lamp install date is in the future (later than the
current instrument time). This can be caused by
changing the system clock to an earlier time or
entering an invalid time when installing the lamp.
17
Invalid pump install date
Pump install date is in the future (later than the
current instrument time). This can be caused by
changing the system clock to an earlier time or
entering an invalid time when installing the pump.
18
Auto-zero failed
Readings not stable enough to complete Auto-zero.
19
TOC value
Invalid TOC warning. TOC reading is above the 1000
ppb limit.