CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018
92 of 112
DLM 97200-03 EN Rev. A
Figure 38: The Washer Inside the Filter Cap
3. Remove the filter element and discard.
4. Remove the new filter element from the packaging.
5. Reassemble the filter. Put the new filter element into the filter (hollow-end first) and put the spring
on top of the element. Place the washer inside the cap. For correct stacking order, see
6. Using the two wrenches, secure the filter cap by turning the wrenches in opposite directions.
Figure 39: Proper Filter Stacking Order
Setting the Installation Date for New Consumables
The
Consumables
screen displays the current state of the lamp, the pump tubing, and (if it is
installed), the battery. From the
Maintenance
menu, select the
Consumables
option.
To install new consumables, press
F1,
and then select
F1
to install a new lamp or
F2
to install a new
pump. The
Horizontal Scroll
bar selects between the day, month and year fields; the
Vertical Scroll
bar changes the value in the selected field.