Portable and mobile RF
communications equipment
should be used no closer to
any part of P
Po
ow
we
err--P
PR
RO
O or
S
SM
MR
RT
T charger, including
cables, than the
recommended separation
distance calculated from the
equation appropriate for the
frequency of the transmitter.
Recommended separation
distance
D=(1.2) (√
P
)
D=(.35) (√
P
)
80 MHz to 800 MHz
D=(0.70) (√
P
)
800 MHz to 2.7 GHz
where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
a
should be less than the
compliance level in each
frequency range.
b
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Conducted RF
IEC 61000-4-6
3 Vrms
6 Vrms in ISM and amateur
radio bands
150kHz to 80MHz
3 V
6 Vrms in ISM and amateur
radio bands
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
N
No
otte
e 1
1:: At 80 MHz and 800 MHz, the higher frequency range applies.
N
No
otte
e 2
2:: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
N
No
otte
e 3
3:: The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio bands
between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz,
10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz, 24.89 MHz to 24.99
MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.
EN
60
6550-309-005 Rev H.0
Содержание Power-PRO TL Cot
Страница 2: ......
Страница 66: ......
Страница 132: ......
Страница 200: ......
Страница 264: ......
Страница 332: ......
Страница 396: ......
Страница 398: ......
Страница 466: ......
Страница 468: ......
Страница 534: ......
Страница 536: ......
Страница 600: ......
Страница 666: ......
Страница 668: ......
Страница 732: ......
Страница 734: ......