BASE CONTROL VERIFICATION
Refer to
1. Optional steer pedal operates properly. Fifth steer wheel operational with steer pedal (green side of the pedal)
engaged.
2. Optional brake pedal operates properly. All casters lock with brake pedal engaged (red side of the pedal).
3. Optional lift pedal operates properly. Litter raises and reaches maximum height (40" - 102 cm) when lift pedal is
pumped.
4. Trendelenburg positions and litter descent are operational when uni-lower pedal (optional) is depressed using
respectively the extremities and its centre.
5. All casters lock and swivel properly.
6. Fixed height stretcher (optional) four casters lock and unlock properly using the brake lever.
7. Litter Movement Verification (see section 2.6):
8. Optional assisted or manual Fowler operates properly.
LITTER MOVEMENT VERIFICATION
Refer to
1. Optional assisted or manual Fowler operates properly.
NOTE: To raise the assisted Fowler, a slight push under the top part of the Fowler is needed to start its rise;
to lower it, push down on its top part while lifting the activation lever.
2. Foot section support arm operates properly.
OTHER EQUIPMENT:
1. Ground chain secured to base.
2. Accessory brackets secure.
3. Optional IV Caddy arm (see
) secure and working properly.
4. Optional two or three-stage IV pole (see
) secured to accessory bracket and is easily
stored on the adjacent accessory bracket.
5. Optional oxygen bottle clamp (see
) secured to hood.
6. Optional chart holder (see
) located under the mattress support secured to frame.
If any problems are found during bed setup, contact the Technical Service department at 1-800-327-0770.