Scale
Operation (Continued)
SCALE OPERATION (CONTINUED)
A. PREPARING (ZEROING) THE SCALE FOR A NEW PATIENT
•
Prepare the stretcher for the new patient (sheets, pillows, equipment, etc.).
•
Press and release the
Weigh/On
button to activate the scale. A weight reading will appear on the display four to ten
seconds later during which a three-hyphen line will flash.
•
Press and keep pressed the
Zero
button during at least two seconds. Release the key; the weight displayed and the
weight stored will be zeroed four to ten seconds later.
•
With the patient in the stretcher, press the
Weigh/On
button. The patient’s weight will be displayed four to ten
seconds later.
CAUTION
To maintain the scale accuracy, it is recommended to zero the scale at least once a week. Note that the patient must be
removed from the stretcher before zeroing the scale.
B. STORING IN MEMORY A WEIGHT READING
•
Press the
Weigh/On
button to activate the scale. Wait for the reading to be displayed.
•
Press and hold the
Memory
button for two seconds. Release the button when the “Memory” message appears on
the display. The weight previously displayed is now stored.
C. RECALLING ON DISPLAY THE WEIGHT STORED IN MEMORY
•
Press the
Weigh/On
button to activate the scale. Do not wait for the weight to be displayed, proceed immediately
with the next step.
•
Press the
Memory
button. The value stored will be displayed along with “Memory”. To leave the
Memory
mode, press
the
Weigh/On
button.
D. ZEROING THE WEIGHT DISPLAYED AND STORED IN MEMORY
•
Press the
Weigh/On
key to activate the scale. Do not wait for the weight to be displayed, proceed immediately with
the next step.
•
Press and hold the
Zero
button for at least two seconds. Release the button; the weight displayed and the weight
stored will be zeroed four to ten seconds later.
E. ZEROING THE MEMORY ONLY
•
Press the
Weigh/On
button to activate the scale. Do not wait for the weight to be displayed, proceed immediately
with the next step.
•
Simultaneously press and hold the
Memory
and
Zero
buttons for at least two seconds and release. The screen will
display 0.0 lb/kg.
F. ADDING OR REMOVING EQUIPMENT WITH A PATIENT IN THE STRETCHER
•
Press the
Weigh/On
button to activate the scale. You may proceed immediately with the next step, but if you want to
compare the weight reading resulting from the
Change Equipment
procedure with the patient’s weight reading before
the procedure, let the weigh process display a weight reading and note the weight.
•
Press the
Change Equipment
button. The “
Change Equipment
” message will appear on the display four to ten
seconds later.
NOTE
: the scale will automatically shut off if this step takes more than three minutes. The operation
will then have to be started over.
•
Add or remove the equipment on the stretcher. Then, press the
Change Equipment
button. The unchanged patient’s
weight will reappear on the screen four to ten seconds later.
NOTE
To prevent erroneous weight readings, the “Adding or Removing Equipment with a Patient in the Stretcher” procedure must be
completed each
time an object is added or removed from the scale weighing zone.
G. CHANGING THE UNIT OF MEASUREMENT
•
Press the
Weigh/On
button to activate the scale. A weight reading will appear on the display four to ten seconds later.
•
Press the
lb /kg
button to change from pounds (lb) to kilograms (kg) or conversely.
•
The
lb /kg
button can be deactivated to prevent inadvertent use of the button. To do so, press and hold the
lb /kg
and
Memory
buttons for at least two seconds. Do the same to reactivate the lb /kg button.