DOC-06800 Rev. A
6–41
Objet1000 - Connex 2 User Guide
3. confirm this in the wizard screen, In the next wizard screen, observe
the numbers and wait until the level is relatively stabletwo units
above or below the average level shown.
Figure 6-54: Level stabilization
4. Select the
Weight is stable
check box, and click
Next
.
If you need to calibrate more than one load cell, the next calibration
screen appears. Repeat steps 3 and 4 until all of the load cells are
calibrated.
5. Click
Done
in the final wizard screen.
Figure 6-55: Final wizard screen