Reprocessing USA and Canada
Instruction For Use • C-MAC
®
Video Laryngoscope 8403xxx and Connection Cable 8403X • 96076019EN-US •
V4.3
33
3. While immersed, use a soft bristle brush (Art. no.: 27652/3) to brush all the surfaces of the
device. Brush for a minimum of 1 minute.
4. Use a low-lint wipe to wipe the entire length of the connecting cable.
5. After the immersion and brushing period, remove the device from the cleaning solution and
rinse using a commercial overhead sprayer nozzle with cold water. Rinse for a minimum of
two minutes.
6. Dry the device and the contacts with a soft, lint-free cloth or dry, oil-free compressed air
(<5 psi).
8.7 Assembly, inspection and care
1. Check the product visually for purity, completeness, damage and dryness.
2. In case of soiling or residue, post-clean the product manually and carry out a complete
cleaning and disinfection process.
3. Sort out damaged and corroded products.
4. Assemble disassembled products.
5. Perform a functional check.
6. For care of the product, refer to the articles in the "Care, Sterilization and Storage
Technology" catalog.
8.8 Packaging
Only use standardized and approved packaging materials and packaging systems (EN 868
Part 2 – 10, EN ISO 11607 Part 1 + 2, DIN 58953).
8.9 Sterilization
The following sterilization method has been validated and approved for this product by KARL
STORZ:
8.9.1 Hydrogen peroxide (H
₂
O
₂
) sterilization - ASP STERRAD
®
WARNING
Risk of infection!
The use of STERRAD
®
processes on lubricated, oiled or pre-assembled products does not
lead to sufficient sterilization.
Sterilize only products that are not oiled and not greased.
Sterilize products in disassembled state.
WARNING
Different STERRAD
®
sterilization systems
Please note that materials and lumen dimensions determine what products can be sterilized
using the various STERRAD
®
sterilization systems.
ATTENTION
Follow the instructions manual for the sterilizer!
The "STERRAD
®
Sterility Guide" can be used to verify whether the relevant medical device
can be sterilized using the different STERRAD
®
devices.
The following STERRAD
®
sterilization methods have been validated and approved for this
medical device by KARL STORZ: