Reprocessing
Instruction For Use • C-MAC
®
Video Laryngoscope 8403xxx and Connection Cable 8403X • 96076019EN-US •
V4.3
29
– STERRAD
®
50, 100S, 200 without booster
– STERRAD
®
NX
®
Standard Cycle
– STERRAD
®
100NX
®
Standard Cycle
7.10.2 Hydrogen peroxide (H
₂
O
₂
) sterilization - STERIS
®
AMSCO
®
V-
PRO™ 1
Detailed information on choosing the appropriate cycles in the various device generations is
available from the manufacturer STERIS
®
.
ATTENTION
Follow the instructions manual for the sterilizer!
WARNING
Risk of infection!
The use of STERRAD V-PRO
®
processes on lubricated, oiled or pre-assembled products does
not lead to sufficient sterilization.
Sterilize only products that are not oiled and not greased.
Sterilize products in disassembled state.
7.10.3 Ethylene oxide sterilization (EO)
The ethylene oxide procedure is validated with 100% ethylene oxide at 55°C and a hold time of
30-45 minutes.
WARNING
Risk of infection!
Application of the EO process to lubricated, oiled or already assembled products does not lead
to sufficient sterilization.
Sterilize only products that are not oiled and not greased.
Sterilize products in disassembled state.
CAUTION
Airing times
For gas sterilization with ethylene oxide, the airing times for instruments stipulated by the
device manufacturer must be observed due to gas absorption of the materials. The airing times
depend on the process engineering of the EO sterilizer (concentration, process control).
7.10.4 Chemical low-temperature sterilization with peracetic acid –
STERIS
®
- System 1
®
For detailed information on the selection of sterilization parameters and on the STERIS
®
Quick
Connect Kit (QCK) required to flush lumina, please contact the manufacturer STERIS
®
.
NOTE: Sterilization is not possible on surfaces which have been lubricated and oiled.
7.10.5 Chemical low-temperature sterilization with peracetic acid –
STERIS
®
System 1E
®
For detailed information on the selection of sterilization parameters and on the STERIS
®
Quick
Connect Kit (QCK) required to flush lumina, please contact the manufacturer STERIS
®
.
Sterilization is not possible on surfaces which have been lubricated and oiled.