Installation and commissioning
Instruction For Use • C-MAC
®
Video Laryngoscope 8403xxx and Connection Cable 8403X • 96076019EN-US •
V4.3
15
4
5
Test for proper operation
Check the instrument for the following characteristics:
– Fully functional control keys
– Securely positioned connecting cable
– Functional image transmission
– Sufficient image quality
– Light transmission
5.2 Initial operation
The C-MAC
®
video laryngoscopes 8403 xxx can be operated with the following monitors:
– C-MAC
®
monitor 8403 ZX (software version: 704v300 or higher)
– C-MAC
®
PM Pocket-Monitor 8403 XD
– with other monitors via the camera control unit C-HUB
®
II 20 2903 01 (software version:
721v105 or higher)
WARNING
Danger due to user error in the wrong viewing mode
Wrong viewing mode
Before each use or after a change of viewing modes/settings, the user should check to
ensure the view observed through the endoscope provides a live image, rather than a
stored one, and has the correct image orientation.
The connecting cable can be removed during operation and connected to another
laryngoscope.
WARNING
Danger due to non-insulated parts
Risk of injury due to electrical current:
It must be ensured that the electrically non-insulated parts of the videoscope do not come
into contact with conductive surfaces, e.g., voltage-carrying elements of other units, at any
time.