Geode H14000 User Manual
11
e) Specifications
General characteristics
Capacity
Up to 12 samples (Sapphire chips) and 48 samples
(Opal chips); (3 chips)/run
Thermo block temperature range
10 °C to 95 °C
Control accuracy thermoblock
± 0.1 °C
Block uniformity (at 72 °C)
± 1.0 °C
Adjustable ramping
from 0.1°C/s to 2°C/s
Pressure input
0 - 1.3 bar
Pressurizing gas
Air
Screen
Touch-sensitive TFT-Display (VGA, Graphics,
65535 colors)
Interfaces
3 x USB, 1 x Ethernet
Dimensions (W x D x H)
35 x 37 x 29 cm
Weight
16 kg ± 0.5 kg
Power supply
100-240V~ // 50/60Hz // 750W
Fuse
T10AH/250V~
Disconnection device
The plug socket serves as a disconnecting device;
the plug socket must be easily accessible.
Maintenance
All parts are required to be examined or supplied
only by the manufacturer or its agent.
Ambient condition
Pollution degree
2
Operation
Environmental conditions
Indoor use only
Temperature
+15 °C to +25 °C
Relative humidity
Max. 70%
Altitude
Operating at max. 1000m above sea level
Noise level
< 37 dB (A)
Storage
Temperature
+5°C to +40°C
Relative humidity
10% to 95% non-condensing
Barometric pressure
700 hPa to 1060 hPa
Waiting time until operation
8h
Shipping
Temperature
-40°C to +60°C
Relative humidity
10% to 95% non-condensing
Barometric pressure
700 hPa to 1060 hPa
Waiting time until the operation
24h
f) Installation of the Geode
i.
Operating requirements
Proper infrastructure requirements for the Geode installation:
o
clean laboratory environment
o
sturdy surface, for supporting from 20 to 40 kg
o
a minimum distance of 25 cm to neighboring objects around the Geode.
o
room temperature between 15°C and 25°C
o
The operating pressure of the Geode should be maintained between 1,250 ± 50 mbars (via the
compressed air system (ISPTLAPG3) delivered with the naica
®
system)
ii.
Installing the Geode:
The initial naica
®
system installation should always be executed by a Stilla Technologies Service
Specialist.
Following the installation, the naica® system performance will be validated according to Stilla
Technologies specifications. At the end of the naica® system installation, Stilla Technologies