23
Command And Response Example Notations
1. [nu]: “not used” at present time, (for future expansion). Undefined data,
data returned should not be used.
2. <Data>: These are ASCII characters shown in Hex notation. The upper
case alpha character set is used for values A through F. (ASCII 46h - F,
ASCII 33h - 3). Lower case is not used.
ASCII Control Character
ASCII in hex
Description
<stx>
02
Start of Text
<etx>
03
End of Text
<ack>
06
Acknowledgment
L
4C
Filter Character
00N
Process Variable Value
100N
Read Set Point Value
200H
Write Set Point Value
Error Messages Reported
Example:Checksum error detected by Instrument at address 32
Instrument responds to Host’s last command with Error message
<stx>
L
<addr> N
<error code>
<ack>
32h N
02h
02
4C 33 32
4E
30
32
06
(Note the absence of Checksum)
N - ASCII character (4Eh) used to denote Error Present
Table Of Error Messages Possible From Instrument
<error code>
00 - Not used.
01 = Undefined command. Command not within acceptable range.
02 = Check-sum error on received data from Host.
03 = Command not performed by Instrument (option not enabled,
restricted
read/write menu, check message)
04 = Illegal ASCII characters received in command. Instrument
accepts only ASCII characters 0 through 9. A through F and a
through f in data field.
05 = Data field error. Not enough, too many, or improper positioning of
characters in data field.
06 = Undefined command. Command not within acceptable range.