STEINDORFF

® 

 

 

 

I.

 

Safety note: 

 

1.

 

Carefully open the box, avoid the accessories, like lens, dropping to ground and being damaged. 

2.

 

Do keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy shaking environment. 

Make sure the stage is smooth, horizontal and firm enough. 

3.

 

When moving the instrument, please use two hands to grip with the two sides of the microscope body. 

4.

 

When running, the lamp house and nearby parts will be very hot. Please ensure there is enough cooling room for 

them. 

5.

 

Always ensure that the grounding terminal of the microscope and that of the wall outlet are properly connected. If 

the proper power cord is not used, product safety performance cannot be warranted. 

6.

 

For safety, be sure the main switch is in “O” (off) state before replace the halogen lamp or the fuse, then cut off the 
power, and do the operation after the lamp bulb and the lamp house completely cool. (Specified Lamp: G6.35, 24v 

100w halogen lamp) 

7.

 

Use the factory supplied power cord, please 

 

II.

 

Maintenance: 

 

1.

 

All the lenses have been well checked and adjusted. It is forbidden to disassemble them yourself. 

2.

 

The nosepiece and coarse/fine focus unit have a compact and precise frame, please don’t disassemble them as 
possible as you can. 

3.

 

Keep the instrument clean, wipe dust regularly, and be attention to avoid contaminating the optical elements 

especially. 

4.

 

The contaminations on the prism, as finger mark and oil, could be gently wiped with a piece of soft cloth or tissue 

paper, gauze which has been immersed in pure alcohol or xylene. (Note that the alcohol and the xylene are all 

burned easily, do not let them near the fire, and use them in a drafty room as possible as you can.) 

5.

 

Don’t use organic solvent to wipe the non-optical elements, when you need to clean, use the soft detergent, please. 

6.

 

When using, if the microscope is splash by liquid, cut off the power at once, and wipe up the moisture. 

7.

 

Do not disassemble any parts of the microscope. That will affect the function or decline the performance of the 

microscope. 

8.

 

Place the instrument in a cool, dry position. After using the microscope, remember to cover it with dust helmet. Do 

wait for the lamp house cooling completely before cover. 

 

III.

 

Electrical Parameters: 

 

1.

 

Voltage Rating: 100-240V AC, 50/60HZ 

2.

 

Power Rating: 200VA 

3.

 

Fuse specs: 50T T5AL250V 

4.

 

Lamp: 24v 100w 

 

 

 

USER NOTICE 

Содержание NYMCS-615

Страница 1: ...lurgical Microscope NYMCS 615 616 It is recommended strongly that you study this manual thoroughly before using the microscope Retain this manual in an easily accessible place near the work desk for f...

Страница 2: ...1 STEINDORFF USER NOTICE 2 1 Component Names 3 2 Assembly 4 3 Operation 9 4 Observation Methods 16 5 Specification 18 6 Troubleshooting 19 CONTENTS...

Страница 3: ...is forbidden to disassemble them yourself 2 The nosepiece and coarse fine focus unit have a compact and precise frame please don t disassemble them as possible as you can 3 Keep the instrument clean...

Страница 4: ...hment optional Mechanical Stage Main Switch Lamp housing Filter Insertion Slot The filter should be inserted from the left Aperture Diaphragm Lever Field Diaphragm Lever Polarizer Insertion Slot Brigh...

Страница 5: ...g be sure every components is clean do not score any parts or glass surface Keep well with the supplied hexagon wrench When changing the components you will need it again 2 INSTALLATION Video Port Tri...

Страница 6: ...h screw down the locking block make the stage support device and the guide board fixed together Fig 3 Fig 1 Fig 2 show the installation sequence of the components which take NYMCS 615 616 for example...

Страница 7: ...device of the mechanical stage reach its low limit position 2 Wresting the lowest magnification objective onto the nosepiece from the left or the right side then push the nosepiece clockwise then plac...

Страница 8: ...until they are against each other turn to a proper position and then screw down the bolt to fix it 2 Insert the power plug into the power jack of the microscope and screw down the bolt to fix it The r...

Страница 9: ...b replacement during or right after use When using or soon after it is turned off the lamp the lamp house and nearby parts will be very hot and will cause serious burns Please turn the main switch on...

Страница 10: ...with the anti direction the voltage decline and the brightness weaken o Using the lamp in a low voltage condition will prolong the use life 3 3 Adjust the Tension Adjustment Collar with Spanner The ti...

Страница 11: ...ng distance objective and want to change another r one be careful of not letting the objective touch the specimen 3 5 Adjusting the Inter pupillary Distance The inter pupillary distance range 48mm 75m...

Страница 12: ...specimen down and search images in the 10 ocular simultaneously then use the fine knob to focus At this moment you can replace other magnification objectives safely and focus without the risk of dest...

Страница 13: ...knob Fig 17 is active Pull the lever out to disengage the ND filter from the light path 3 9 Using the Filters o Engage the optimum filter sliders for the purpose of observation in the two filter inse...

Страница 14: ...s diaphragm a little 4 Rotate the two FS centering screws using the Allen screwdriver to adjust so that the field iris image becomes concentric with the field of view 5 While pushing in the FS knob op...

Страница 15: ...ure number of the objective fig 29 In reflected light darkfield observation the aperture must be fully opened by pushing in the AS knob o With some specimens an image with high contrast and little fla...

Страница 16: ...iaphragm The aperture diaphragm is designed for the adjustment of the numerical aperture not for the brightness Generally reducing the diaphragm opening to 70 80 of the N A value of the respective obj...

Страница 17: ...ilter insertion slot Brightness adjustment knob P 13 P 9 P 12 P 13 P 10 P 11 P 9 P 10 P 10 P 14 P 13 P 12 P 9 Start observation Adjust the brightness Insert the required filters Engage the desired obj...

Страница 18: ...ounting knob on the front of the DIC revolving nosepiece insert the DIC slider and clamp by tightening the mounting knob 2 Observation method 2 1 Place the specimen on the stage and adjust the focus b...

Страница 19: ...lumination Aspheric collector Polarizer and analyzer Coupling plate Blue green yellow filter and frosted glass Transmitted light illumination Swing out condenser NA0 9 0 25 24v 100w halogen bulb Asphe...

Страница 20: ...e knob correctly Revolving nosepiece is not in a click position Set it in a click position Field iris diaphragm is not centered Center the field iris diaphragm correctly and open it sufficiently ND fi...

Страница 21: ...distance is incorrect Adjust it again The eyes overtire Incorrect diopter adjustment Adjust diopter correctly Brightness uncomfortable Adjust the bulb voltage 6 3 Electric Part PROBLEM REASON FOR PROB...