5
STEINDORFF
®
2.3.
Installation Steps
2.2.1.
Installing the Mechanical Stage Support Device
1.
Before installing the device, be sure to adjust the coarse
focus knob
①
. Make the guide board
②
down to the lowest
position, so you can install the mechanical stage support
device easily.
2.
Hold on the mechanical stage support device (Fig. 2), place
it from the top of the guide board (Fig. 1), and let the device
(Fig. 2) falling free until it reach the limit bolt. Use the
hexagon wrench screw down the locking block
③
, make the
stage support device and the guide board fixed together (Fig.
3).
Fig. 1, Fig. 2 show the installation sequence of the
components which take NYMCS-615/616 for example.
For BS-6022RF, there are two limit bolt holes with
different height on the guide board
②
to meet with
sample which have different thickness: the upside hole is
for sample with 0
~
30mm thickness, the underside hole
is for sample with 20
~
50mm thickness.
The mechanical stage
⑤
have been adjusted horizontally
and fixed together before leaving factory. Do not
disassemble unless necessary, that may affect the
observation precision of the instrument.
①
②
Limit
bolt
④
⑤
⑤
Fig. 1
Fig. 2