background image

14 

STEINDORFF

® 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig. 29 

Fig. 30 

Fig. 31 

The image of aperture iris diaphragm 

Polarizer 

 

Analyzer 

Coupling plate 

 

 

 

3.12.

 

Centering the Aperture Iris Diaphragm (AS) 

1.

 

Slide the mirror selector lever to “BF”. 

2.

 

Engage the 10X objective by rotating the revolving 

nosepiece, place the specimen on the stage and adjust 

approximate focusing. 

3.

 

Remove the eyepiece, look into the eyepiece sleeve and pull 

the AS knob 

, so that the aperture is about 70%. 

4.

 

If the center of the iris diaphragm is deviated, center it by 

rotating the two AS centering screws 

 (Fig. 27) using the 

Allen screwdriver. 

 

Using the Aperture Iris Diaphragm 

 

In reflected light brightfield observation, optimum 

observation is generally possible by setting the aperture to 

between 70% and 80% of the aperture number of the 

objective (fig. 29). 

 

In reflected light darkfield observation, the aperture must be 

fully opened by pushing in the AS knob. 

 

o

 

With some specimens, an image with high contrast and little 

flare may sometimes be obtained when the aperture is 

slightly closed. It is therefore recommended to also try a 

slightly closed aperture. 

 

3.13.

 

Setting the Analyzer and Polarizer 

1.

 

Insert the polarizer (Fig. 30) into the polarizer insertion Slot 

 with the surface printing with Silk Screen towards you, 

then push the polarizer into the light path. 

2.

 

Remove the cover, then put the analyzer (Fig. 30) in the 

insertion slot 

 (Fig. 31). 

3.

 

Rotate the analyzer rotating dial 

 to find the position 

where the field of view is darkest. 

4.

 

When the analyzer and polarizer are coupled by using the 

coupling plate | provided with the polarizer and tightening the 

clamping knobs on it, the analyzer and polarizer can be 

engaged or disengaged in the light path together (Fig. 31) 

Содержание NYMCS-615

Страница 1: ...lurgical Microscope NYMCS 615 616 It is recommended strongly that you study this manual thoroughly before using the microscope Retain this manual in an easily accessible place near the work desk for f...

Страница 2: ...1 STEINDORFF USER NOTICE 2 1 Component Names 3 2 Assembly 4 3 Operation 9 4 Observation Methods 16 5 Specification 18 6 Troubleshooting 19 CONTENTS...

Страница 3: ...is forbidden to disassemble them yourself 2 The nosepiece and coarse fine focus unit have a compact and precise frame please don t disassemble them as possible as you can 3 Keep the instrument clean...

Страница 4: ...hment optional Mechanical Stage Main Switch Lamp housing Filter Insertion Slot The filter should be inserted from the left Aperture Diaphragm Lever Field Diaphragm Lever Polarizer Insertion Slot Brigh...

Страница 5: ...g be sure every components is clean do not score any parts or glass surface Keep well with the supplied hexagon wrench When changing the components you will need it again 2 INSTALLATION Video Port Tri...

Страница 6: ...h screw down the locking block make the stage support device and the guide board fixed together Fig 3 Fig 1 Fig 2 show the installation sequence of the components which take NYMCS 615 616 for example...

Страница 7: ...device of the mechanical stage reach its low limit position 2 Wresting the lowest magnification objective onto the nosepiece from the left or the right side then push the nosepiece clockwise then plac...

Страница 8: ...until they are against each other turn to a proper position and then screw down the bolt to fix it 2 Insert the power plug into the power jack of the microscope and screw down the bolt to fix it The r...

Страница 9: ...b replacement during or right after use When using or soon after it is turned off the lamp the lamp house and nearby parts will be very hot and will cause serious burns Please turn the main switch on...

Страница 10: ...with the anti direction the voltage decline and the brightness weaken o Using the lamp in a low voltage condition will prolong the use life 3 3 Adjust the Tension Adjustment Collar with Spanner The ti...

Страница 11: ...ng distance objective and want to change another r one be careful of not letting the objective touch the specimen 3 5 Adjusting the Inter pupillary Distance The inter pupillary distance range 48mm 75m...

Страница 12: ...specimen down and search images in the 10 ocular simultaneously then use the fine knob to focus At this moment you can replace other magnification objectives safely and focus without the risk of dest...

Страница 13: ...knob Fig 17 is active Pull the lever out to disengage the ND filter from the light path 3 9 Using the Filters o Engage the optimum filter sliders for the purpose of observation in the two filter inse...

Страница 14: ...s diaphragm a little 4 Rotate the two FS centering screws using the Allen screwdriver to adjust so that the field iris image becomes concentric with the field of view 5 While pushing in the FS knob op...

Страница 15: ...ure number of the objective fig 29 In reflected light darkfield observation the aperture must be fully opened by pushing in the AS knob o With some specimens an image with high contrast and little fla...

Страница 16: ...iaphragm The aperture diaphragm is designed for the adjustment of the numerical aperture not for the brightness Generally reducing the diaphragm opening to 70 80 of the N A value of the respective obj...

Страница 17: ...ilter insertion slot Brightness adjustment knob P 13 P 9 P 12 P 13 P 10 P 11 P 9 P 10 P 10 P 14 P 13 P 12 P 9 Start observation Adjust the brightness Insert the required filters Engage the desired obj...

Страница 18: ...ounting knob on the front of the DIC revolving nosepiece insert the DIC slider and clamp by tightening the mounting knob 2 Observation method 2 1 Place the specimen on the stage and adjust the focus b...

Страница 19: ...lumination Aspheric collector Polarizer and analyzer Coupling plate Blue green yellow filter and frosted glass Transmitted light illumination Swing out condenser NA0 9 0 25 24v 100w halogen bulb Asphe...

Страница 20: ...e knob correctly Revolving nosepiece is not in a click position Set it in a click position Field iris diaphragm is not centered Center the field iris diaphragm correctly and open it sufficiently ND fi...

Страница 21: ...distance is incorrect Adjust it again The eyes overtire Incorrect diopter adjustment Adjust diopter correctly Brightness uncomfortable Adjust the bulb voltage 6 3 Electric Part PROBLEM REASON FOR PROB...

Отзывы: