STEINDORFF

® 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.2.8.

 

Attaching the Halogen Bulb 

 

The applicable lamp bulb is the 24v 100w Halogen. 

1.

 

Fully loosen the clamping screw 

 at the top of the lamp 

housing using the provided hexagon wrench. 

2.

 

Remove the lamp housing 

 by lifting it up. 

3.

 

Tilt the bulb socket by 90° (Fig. 14) 

4.

 

While pushing down the bulb clamping levers, hold the 

halogen bulb with gloves or a piece of gauze, insert the bulb 

pins into the sections as far as they will go. Then return the 

lamp clamping lever gently back to the original position to 

clamp the bulb. 

 

Caution for bulb replacement during or right after use 

When using, or soon after it is turned off the lamp, the lamp house and 

nearby parts will be very hot and will cause serious burns. Please turn 

the main switch on “O” (off), pull out power plug, and make sure the 

bulb, the lamp room and periphery are all cool. Then, you can do your 

replacing. 

 

 

Please insert the lamp gently, or it will be damaged by 

excessive extrusion. 

 

 

Do not touch the Halogen bulb with your bare hands. It 

will shorten the service life or cause it to burst. If you 

leave finger marks on the surface carelessly, clean it with 

a dry soft cloth. 

 

 

When using, the temperature of the lamp housing surface 

will be very hot, please pay attention to the warning 

written on the Warning board. 

Fig. 13 

Fig. 14 

Fig. 15 

 

 

 

 

Содержание NYMCS-615

Страница 1: ...lurgical Microscope NYMCS 615 616 It is recommended strongly that you study this manual thoroughly before using the microscope Retain this manual in an easily accessible place near the work desk for f...

Страница 2: ...1 STEINDORFF USER NOTICE 2 1 Component Names 3 2 Assembly 4 3 Operation 9 4 Observation Methods 16 5 Specification 18 6 Troubleshooting 19 CONTENTS...

Страница 3: ...is forbidden to disassemble them yourself 2 The nosepiece and coarse fine focus unit have a compact and precise frame please don t disassemble them as possible as you can 3 Keep the instrument clean...

Страница 4: ...hment optional Mechanical Stage Main Switch Lamp housing Filter Insertion Slot The filter should be inserted from the left Aperture Diaphragm Lever Field Diaphragm Lever Polarizer Insertion Slot Brigh...

Страница 5: ...g be sure every components is clean do not score any parts or glass surface Keep well with the supplied hexagon wrench When changing the components you will need it again 2 INSTALLATION Video Port Tri...

Страница 6: ...h screw down the locking block make the stage support device and the guide board fixed together Fig 3 Fig 1 Fig 2 show the installation sequence of the components which take NYMCS 615 616 for example...

Страница 7: ...device of the mechanical stage reach its low limit position 2 Wresting the lowest magnification objective onto the nosepiece from the left or the right side then push the nosepiece clockwise then plac...

Страница 8: ...until they are against each other turn to a proper position and then screw down the bolt to fix it 2 Insert the power plug into the power jack of the microscope and screw down the bolt to fix it The r...

Страница 9: ...b replacement during or right after use When using or soon after it is turned off the lamp the lamp house and nearby parts will be very hot and will cause serious burns Please turn the main switch on...

Страница 10: ...with the anti direction the voltage decline and the brightness weaken o Using the lamp in a low voltage condition will prolong the use life 3 3 Adjust the Tension Adjustment Collar with Spanner The ti...

Страница 11: ...ng distance objective and want to change another r one be careful of not letting the objective touch the specimen 3 5 Adjusting the Inter pupillary Distance The inter pupillary distance range 48mm 75m...

Страница 12: ...specimen down and search images in the 10 ocular simultaneously then use the fine knob to focus At this moment you can replace other magnification objectives safely and focus without the risk of dest...

Страница 13: ...knob Fig 17 is active Pull the lever out to disengage the ND filter from the light path 3 9 Using the Filters o Engage the optimum filter sliders for the purpose of observation in the two filter inse...

Страница 14: ...s diaphragm a little 4 Rotate the two FS centering screws using the Allen screwdriver to adjust so that the field iris image becomes concentric with the field of view 5 While pushing in the FS knob op...

Страница 15: ...ure number of the objective fig 29 In reflected light darkfield observation the aperture must be fully opened by pushing in the AS knob o With some specimens an image with high contrast and little fla...

Страница 16: ...iaphragm The aperture diaphragm is designed for the adjustment of the numerical aperture not for the brightness Generally reducing the diaphragm opening to 70 80 of the N A value of the respective obj...

Страница 17: ...ilter insertion slot Brightness adjustment knob P 13 P 9 P 12 P 13 P 10 P 11 P 9 P 10 P 10 P 14 P 13 P 12 P 9 Start observation Adjust the brightness Insert the required filters Engage the desired obj...

Страница 18: ...ounting knob on the front of the DIC revolving nosepiece insert the DIC slider and clamp by tightening the mounting knob 2 Observation method 2 1 Place the specimen on the stage and adjust the focus b...

Страница 19: ...lumination Aspheric collector Polarizer and analyzer Coupling plate Blue green yellow filter and frosted glass Transmitted light illumination Swing out condenser NA0 9 0 25 24v 100w halogen bulb Asphe...

Страница 20: ...e knob correctly Revolving nosepiece is not in a click position Set it in a click position Field iris diaphragm is not centered Center the field iris diaphragm correctly and open it sufficiently ND fi...

Страница 21: ...distance is incorrect Adjust it again The eyes overtire Incorrect diopter adjustment Adjust diopter correctly Brightness uncomfortable Adjust the bulb voltage 6 3 Electric Part PROBLEM REASON FOR PROB...

Отзывы: