4.8 Shutting off the Alarm Volume
Set the
Min.Alm.Vol.
and Alarm Vol.
to 0 to shut off the alarm volume. Then
there will be a symbol shown in the alarm status area. The alarm lamp and
alarm messages are still active after the alarm volume is off. The audible alarm is
reactivated automatically when:
• The factory configuration is loaded;
• Set the alarm volume to a nonzero value.
When a factory configuration is selected, the alarm volume of the monitor
may be lower than the minimum alarm volume. In this case the alarm volume is
automatically adjusted according to the minimum alarm volume.
Warning:
•
When the alarm sound is switched off, the monitor will
give no audible alarm tones even if a new alarm occurs.
Therefore the user should be very carefully about weather
to switch off the alarm sound or not.
•
Don’t rely exclusively on the audible alarm system for
patient monitoring. Adjusting alarm volume to a low level
may result in a hazard to the patient. Always keep the
patient under close surveillance.
4.9 When an Alarm Occurs
Note: When an alarm occurs, you should always check the
patient’s condition first.
Check the alarm message appeared on the screen. It is needed to identify the
alarm and action appropriately, according to the cause of the alarm.
1. Check the patient’s condition.
2. Identify alarming parameter and alarm category.
3. Identify the cause of the alarm.
4. Silence the alarm, if necessary.
5. When cause of alarm has been over, check that the alarm system
is working properly.
You will find the alarm messages for the individual parameter in
Appendix D
Alarm message.