88
5.6.3 Adjusting the Alarm Limit
1. Select
Menu
Alarm Setup
.
2. Adjust
High
:If the SpO
2
measurement is higher than the high alarm limit, the
“SpO
2
Too High” alarm will be triggered.
3. Adjust
Low
:If the SpO
2
measurement is lower than the low alarm limit, the
“SpO
2
Too Low” alarm will be triggered.
5.6.4 Setting Desat Limit
SpO
2
desat means when SpO
2
measuring value is lower than the desat limit, a
high physiological alarm will be trigged. Its setting is as follows:
1. Select
Menu
System
Maintenance
,then pops up a password entering
window.
2. Input the password and select
OK
to enter the maintenance window. Select
Desat Limit
, Then set its value through the Up and Down button.