8.3 Cleaning the Monitor
1. Common detergent and non-corrosive disinfectant used in hospital can be
applied to clean monitor, however you must be aware that many kinds of
detergents must be diluted prior to utilization, and please use it according to
the instruction of detergent manufacturer.
2. Avoid the use of alcohols, amino or acetonyl detergent.
3. The enclosure and screen of monitor shall be free of dust, and they can be
wiped with lint-free soft cloth or sponge soaked in detergent. While cleaning,
be careful and do not spill liquid onto the instrument and keep any liquid out
of it. When wiping the side panel of monitor, you must be especially careful to
keep water out of all kinds of cable and outlet on the panel.
4. Do not use abrasive material including wire brush or metal brightener during
cleaning because this material will damage the panel and monitor screen.
5. Do not submerge the monitor in liquid.
6. While cable or plug of attachment accidentally gets wet, please rinse it with
distilled water or deionized water and dry it in the environment of temperature
40°C to 80°C for at least one hour.
8.4 Cleaning SpO
2
Sensor
1. The casing of the sensor and light tube can be cleaned with swab or non-velvet
soft cloth dipped with medical alcohol.
2. The sensor cable can be cleaned or sterilized with Hydrogen Peroxide 3% or
isopropyl alcohol 70%.
3. It is forbidden to put the monitor in high-pressure containers and put the
sensor directly in liquid.
Warning: Do not reuse or disinfect the disposable SpO
2
sensor.
8.5 Disposal
Dispose of the monitor in accordance with local environment and waste disposal
laws and regulations. For the disposal of SpO
2
sensor, follow local regulations
regarding disposal of hospital waste.