Immunity
test
IEC
60601
test level
Com-
pliance
level
Electromagnetic environment -
guidance
Conducted
RF
IEC 61000-
4-6
Radiated
RF IEC
61000-4-3
3 Vrms
150 kHz
to 80
MHz
3 V/m
80 MHz
to 2.5
GHz
1 V rms
3 V/m
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the MASTER
PALM 3® Handheld pulse oximeter, in-
cluding cables, than the recommended
separation distance calculated from
the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance
p
1
5
.
3
=
v
d
P
E
d
=
1
5
.
3
80 MHz-800 MHz
P
E
d
=
1
7
800 MHz-2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recom-
mended separation distance in metres
(m).
Field strengths from fixed RF transmit-
ters, as determined by an electro-
magnetic site survey,a should be less
than the compliance level in each
frequency range.b
Interference may occur in the vicinity
of equipment marked with the fol-
lowing symbol: