3.6.2 Spot-checking Mode
Spot-checking mode is intended for short-term on-site measurement. This mode
is normally selected to check outpatient when doctors make rounds of the
wards. The patient ID will automatically increase from 1 to 99 according to the
connecting of SpO
2
sensor. Details are as follows:
Apply the SpO
2
sensor to the patient.
After valid signals are detected,
1. The patient ID flashes and automatically increases by 1 after 8 seconds to admit
a new patient.
2. Press the Left button when the current patient ID is flashing, the patient ID will
stop flashing and remain unchanged. The patient will not be admitted and new
measurements will be stored under the current patient ID.
3. When the storage of patient measuring data reaches its limit, the newly
measuring data will cover for the primary one.
3.7 Selecting Patient Type
To select the patient type,
1. Select
Menu
System
Type
.
2. Set
Type
to Adu
(adult), Ped
(child) or Neo
(neonate).
3.8 Entering/Exiting the Demo Mode
To enter the demo mode:
1. Select
Menu
System
Maintenance
enter the required password.
2. Set
Screen
to
Demo
and the message
Demo Mode
is shown in the
technical alarm area.
To exit the demo mode:
1. Select
Menu
System
Maintenance
enter the required password.
2. Set
Screen
to
Normal
.
Caution: Only when the monitor isn’t monitoring any patient,
connecting its SpO2 sensor to a patient, the patient ID will
add 1 automatically.