Modbus Communication
OXY5500 Operator’s Manual
4
–
5
Table 4-1
Holding registers (Continued)
Register
Name
Address
Size
Variable
Type
Description
Write
Access
Device ID
RS-485
4095
2
Int32
Set the device ID that is
used in Modbus RTU com-
munication (range 1-32).
Refer to Table 4–6.
Yes
Device ID
Minimum
RS-485
4097
2
Int32
Device ID Address Limit:
Minimum.
No
Device ID
Maximum
RS-485
4099
2
Int32
Device ID Address Limit:
Maximum.
No
Baud rate
RS-485
4101
2
Int32
Code for Baud rate where:
3 = 9600
4 = 19200
5 = 38400
6 = 57600
7 = 115200
Yes
Baud rate
Minimum
RS-485
4103
2
Int32
Minimum code for baud
rate.
No
Baud rate
Maximum
RS-485
4105
2
Int32
Maximum code for baud
rate.
No
Parity
RS-485
4107
2
Int32
Parity for the RS-485 out-
put where:
0x00 = Even Parity
0x01 = Odd Parity
0x02 = No Parity
1
1. When using “No Parity,” an additional Stop-bit will be sent (2 Stop bits).
Yes
Device ID
RS-232
4109
2
Int32
Sets the device ID that is
used in the Modbus RTU
communication (range 1 -
32). Refer to Table 4–7.
Yes
Device ID
Minimum
RS-232
4111
2
Int32
Device ID Address Limit:
Minimum.
No
Device ID
Maximum
RS-232
4113
2
Int32
Device ID Address Limit:
Maximum.
No
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...