Operation
OXY5500 Operator’s Manual
3
–
17
2.
Change the
O2-2nd
value according to the value shown below the
3.
Click the
Save
button in the upper right of the screen to store the
changes and complete the manual calibration of the analyzer.
The display will automatically change to the MEASUREMENT window.
If another sensor type has been selected, a message window will
display stating that the sensor type change has reset RATA (refer to
“Relative Accuracy Test Audit (RATA)”
on page 3-22). Refer to
Two-point calibration
Use the following procedure to perform a two-point calibration with the
connected oxygen sensor.
Confirm that the correct unit for the
O2-2nd
and
pATM
values
are selected.
Figure 3–21
Message window - Sensor type
change resets RATA