Operation
OXY5500 Operator’s Manual
3
–
49
4-20mA calibration
Use the 4-20mA CALIBRATION screen to calibrate the
Output
and
Input
. The
analyzer is delivered in the calibrated state, but can be calibrated to other
devices in your measurement system.
Output calibration
Use the following procedure to calibrate the work flow for either Output 1 or
Output 2. Refer to Figure 3–55.
1.
Connect a current measurement device to the respective output.
This will serve as the reference device.
2.
Set the
1st Point
value to any low value; e.g., 4.00mA. The value
will be applied immediately.
a.
Press the
Apply
button or enter another value.
3.
Read the current value shown on the reference device; e.g., 3.90mA.
a.
Use the +/- symbols in the
Adjust
column next to the
1st
Point
value to adjust the values accordingly.
Factory calibration will be lost if the analyzer is re-calibrated.
Figure 3–54
Analogues - Concentration alarm
relay
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...