Modbus Communication
OXY5500 Operator’s Manual
4
–
7
Table 4-1
Holding registers (Continued)
Register
Name
Address
Size
Variable
Type
Description
Write
Access
4-20mA
Port1 Fixed
Value
4383
2
Float
In fixed output mode, this
value is applied to the out-
put. Unit is mA.
Yes
4-20mA
Port1 Error
Trigger Level
Value
4389
2
Int32
Output current in the event
of an error, where:
0x00 = 22mA
0x01 = 2mA
Yes
4-20mA
Port1 Calibra-
tion Values
4329
8
Float
2 calibration values for a
low point and a high point
(each with a reference
value and device output).
Refer to
4-20mA
Port2 Output
Interface
4945
2
Int32
Code for 4-20mA Port1
Output Mode where:
0x00 = off
0x01 = fixed
0x02 = linear
0x04 = bilinear
Yes
4-20mA
Port2 Output
Channel
4949
2
Int32
Code for 4-20mA Port1
Output Interface where:
0x01 = Oxygen
0x20 = Temperature
Yes
4-20mA
Port2 Low
Value
4963
2
Float
The 4mA output value.
Refer to
Yes
4-20mA
Port2 Mid
Value
4965
2
Float
The 12mA output value is
only used in bilinear mode.
Refer to
Yes
4-20mA
Port2 High
Value
4967
2
Float
The 20mA output value.
Refer to
Yes
4-20mA
Port2 Fixed
Value
4969
2
Float
In fixed output mode, this
value is applied to the out-
put.
Yes
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...