Oxygen Analyzer
3
–
50
4900002239 rev. G 7-11-17
4.
Set the
2nd Point
value to any high value; e.g., 20.00 mA. The
value will be applied immediately.
a.
Press the
Apply
button or enter another value.
5.
Read the current value shown on the reference device; e.g.,
19.54mA.
a.
Use the +/- symbols in the
Adjust
column next to the
2nd
Point
value to adjust the values accordingly.
Example
: The analyzer shows a value of 19.54mA and the value should be
20.00mA. Press the “+” button until the desired value is
registered.
To test the calibration, apply some test points by selecting different percentage
values, such as 0%, 25%, 50%, 75% or 100%, which corresponds to 4mA,
8mA, 12mA, 16mA and 20mA. Check the values with the reference device. If
satisfied with the calibration, press the
button to save.
Input calibration
The procedure for calibrating the input is similar to the output procedure noted
above. Use the following steps to calibrate the input. Refer to Figure 3–56.
1.
Apply a low current to the OXY5500.
2.
Enter this value in the
Reference
column in the
1st Point
row.
Figure 3–55
Analogues - 4-20mA calibration
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...