Operation
OXY5500 Operator’s Manual
3
–
51
3.
Press the
Set
button next to the
1st Point
when the reading is
steady. The last measured value will be displayed in the top row next
to the selected port.
4.
Apply a higher value to the OXY5500.
5.
Enter this value in the
Reference
column in the
2nd Point
row.
6.
Press the
Set
button next to the
2nd Point
when the reading is
steady.
7.
The row
Test Point
displays the calibrated value, which is used for
calculating the pressure value. This value should match the
reference device value by less than 0.05mA.
This value is the uncalibrated value that will be used as the
1st
Point
calibration value.
This value is the uncalibrated value that will be used as the
2nd
Point
calibration value.
Figure 3–56
Analogues - 4-20mA input
calibration
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...