4-20mA interface settings
The 4-20mA interface settings can be accessed through the ANALOGUES
menu. The following screen displays (refer to Figure 3–49).
The
Output
,
Mode
and
Error Trigger Level
settings will be applied to the
selected
Port
, which includes Port1, Port2 or the Input.
The
Output
of Port1 or Port2 can be
Oxygen
or
Temperature
.
The
Input
is always
Pressure
and cannot be changed.
The
Mode
of Port1 and Port2 can be set to one of the following:
•
Off
: No input reading or output writing.
•
Linear
: High and low value that has been set to correspond to 4mA
and 20ma. Values between these two settings will be calculated
linearly. Values outside this range will initiate the error trigger level.
•
Bilinear
: High, mid and low value set to correspond to 4mA, 12mA
and 20mA respectively. This mode allows higher resolution in a
certain range. Refer to Figure 3–50 for an example.
Figure 3–49
Analogues - 4-20mA interface
settings
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...