Oxygen Analyzer
4
–
18
4900002239 rev. G 7-11-17
Ethernet Subnet Mask
Examples
Configuration of a Continuous Measurement
Precondition
: Sensor is connected, sensor constants and calibration values
are already set up correctly (OP-9).
The goal for this configuration is a continuous measurement with a 1 minute
interval, pressure sensor and RTD (Pt100) are deactivated. Instead a manual
fixed value will be transmitted. Refer to Table 4–17.
Table 4–16
Definition of register 5685
Start
Register
Number of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Write
Access
5685
8
Int32:
Ethernet
Subnet Mask
Octet 1
Int32:
Ethernet
Subnet
Mask
Octet 2
Int32:
Ethernet
Subnet
Mask
Octet 3
Int32:
Ethernet
Subnet
Mask
Octet 4
Yes
Table 4–17
Configuration for a continuous measurement
Step
Description
Register(s)
Value
1
Stop the measurement, in case it is
running.
5707, 5708
0 (Int32)
2
Set pressure mode to ‘Manual’.
5705
0 (Boolean)
3
Set manual pressure to “1006.23”.
3147, 3148
1006.23 (Float)
4
Set temperature mode to “Manual”. 5611
0 (Boolean)
5
Set manual temperature to
“20.56”.
2409, 2410
20.56 (Float)
6
Set the interval rate to 1 min.
(60s).
3499, 3500
60 (Int32)
7
Activate the interval mode and
start the continuous measurement
immediately.
5707, 5708
5 (Int32 trans-
lates to 00000101
binary)
Содержание OXY5500
Страница 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Страница 2: ......
Страница 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...