Section 4: Software Commands
NetClock/GTP Instruction Manual
Page 4-7
4.5
RESTORE FACTORY DEFAULTS
The
DEF
command returns all selectable parameters to the factory default settings.
The clock must be placed in
Test Mode
to execute the
DEF
command. The factory
default settings are listed below:
FRONT PANEL FORMAT= 24 HOUR
TIME DIFF= +00:00 DST= 0
IRIG FORMAT= FAA
TIME DIFF= +00:00 DST= 0
SERIAL PORT 1
BAUD RATE= 9600 FORMAT #= 00 REQUEST CHAR= T
TIME DIFF= +00:00 DST= 0
SERIAL PORT 2
BAUD RATE= 9600 FORMAT #= 00 REQUEST CHAR= T
TIME DIFF= +00:00 DST= 0
REMOTE PORT 1
BAUD RATE= 9600 FORMAT #= 00
TIME DIFF= +00:00 DST= 0
LOCK TIME OUT= 00:15:00
SYNC TIME OUT= 02:00:00
To restore the clock to factory default settings enter the following:
Type:
TM ON <ent>
Sample
response:
TIME= 15:31:57
DATE= 2001-04-18 STATUS CHANGE
ACTIVE ALARMS: MAJOR AND MINOR
IN TEST MODE
Type:
DEF <ent>
Response:
ALL PARAMETERS SET TO FACTORY DEFAULTS
Содержание 8183A GTP
Страница 36: ...Section 3 Operation Page 3 2 NetClock GTP Instruction Manual FIGURE 3 1 NETCLOCK GTP FRONT PANEL...
Страница 39: ...Section 3 Operation NetClock GTP Instruction Manual Page 3 5 FIGURE 3 2 NETCLOCK GTP REAR PANEL...
Страница 64: ...SECTION 4 SOFTWARE COMMANDS 4 0 INTRODUCTION 4 1 4 22 RS 232 COMMAND SET...
Страница 96: ...SECTION 5 SERVICE INFORMATION 5 0 INTRODUCTION 5 1 RECEPTION TROUBLESHOOTING 5 2 TCXO ADJUSTMENT...
Страница 102: ...APPENDIX A IRIG CODE DESCRIPTION A 0 INTRODUCTION A 1 FAA IRIG B CODE...
Страница 104: ...Appendix A FAA IRIG Code Description Page A 2 NetClock GTP Instruction Manual FIGURE A 1 FAA IRIG B TIME CODE...