Manual V2.docx | Status 10.12.2015 | Version 1.0
5.
In the SpinDetectorControl device, click on
Set Voltages
, see Figure 44. Set the
parameter values from the SPECS Test Report as start values. Typical voltages are
shown in Table 11. Set the voltages and click
Apply
.
Figure 41. Screenshot of the Device Controls window for the Spin Detector Control in SpecsLab Prodigy.
Table 11. Typical parameter values for the Ferrum Detector when operated in total intensity mode.
Parameter
Value (V)
Scattering Energy
400
Lens 1
200
Lens 2
50
Lens 3
400
Lens 4
400
Lens1DeflectorX
-1.2
Lens1DeflectorY
0
Lens1Stigmator
0
Lens3DeflectorX
-8.5
Lens3DeflectorX
0
Lens3Stigmator
0
6.
Optimize the intensity by adjusting spin deflector 1 and 2, i.e., adjust SDefl1X,
SDefl1Y, SDefl2X and SDefl2Y in the PHOIBOS 1D
Para. Voltages
, see Figure 42. It is
necessary to click
Apply
after each change.