Appendix C: Troubleshooting
FTS Systems FlexiCool
28
Rev 001, 10/13
© SP Scientific 2013
Symptom
Possible Cause
Remedy
Unit will not control and/or the
Process Value (PV) Display on the
SOLO controller is flashing
PV is above the upper span limit
Wait several minutes as the
system cools. The condition will
self- correct.
PV is below the lower span limit
Turn off the main breaker to permit
the system to warm naturally.
External Pencil Sensor (RTD) not
connected
Verify that the external pencil
sensor (RTD) is connected to the
jack labeled SENSOR on the back
of the unit and that it is in contact
with the fluid / vapor that is being
controlled.
Note:
For information on span limits, refer to Appendix B: Changing