Air Flow Considerations
The FTS Systems FC50 and FC100 are equipped with an air-cooled refrigeration
system. Air is pulled from the left side of the unit to cool the refrigeration system
components and then exits the rear panel. When positioning your system, ensure that
it is located on a firm, level surface in an area that provides adequate air circulation.
Inadequate airflow will degrade the unit’s cooling capacity and in extreme cases may
result in compressor failure. A minimum of 4 inches (10.2 cm) of clearance on all
sides of the unit is required to ensure proper airflow and avoid damage to the
refrigeration system.
This equipment can add up to 3400 Btu/hr (1 kw) of heat to the room. Always ensure
that proper ambient temperature is maintained to ensure best performance.
Electrical Requirements
All 120 VAC / 60 Hz equipment is supplied with a 6 ft., 7 in (2 m) long power cord
terminated with a molded NEMA 5-15P plug.
All 220 VAC / 50 Hz equipment is supplied with a 9 ft., 10 in (3 m) long, un-terminated,
AWG 18-3 power cord. Both cords are detachable and have a molded IEC 60320
C-13 plug that attach to the machine.
Equipment rated with a nominal voltage of 120 VAC / 60 Hz should only be operated
from 113 to 127 VAC. Equipment rated with a nominal voltage of 220 VAC / 50 Hz
should only be operated from 211 to 240 VAC.
Note:
Variances in frequency of ± 3 Hz are permitted.
IEC Protection - Class 1
This unit’s chassis is connected to earth ground through a 3-conductor mains cable.
Ensure that the AC outlet has corresponding ground and that the voltage and
frequency match the serial tag on the back of the unit.
Services and Utilities
The FTS Systems FlexiCool’s serial tag, which is located on the unit’s back panel,
provides the unit serial number, and electrical requirements. Ensure that the voltage,
phase, frequency and amperage listed on the serial tag match your facility’s available
power supply (
e.g.,
as a minimum the power outlet you intend to use must meet the
requirements listed on the serial tag).
CAUTION! IF YOU ARE UNSURE ABOUT THE AVAILABLE ELECTRICAL
VOLTAGE SUPPLY IN YOUR FACILITY, CONSULT A QUALIFIED ELECTRICIAN.
Prior to connecting your FTS Systems FlexiCool to the available electrical supply,
ensure that the Main Power switch is in the Off position. You may connect the unit to
the available electrical supply at this time.
NEMA 5-15P
Содержание FTS FLEXICOOL IMMERSION COOLERS FC100
Страница 1: ...IMMERSION COOLERS FC50 FC100 OPERATOR S MANUAL FTS SYSTEMS FLEXICOOL IMMERSION COOLERS ...
Страница 2: ......
Страница 6: ...iv Rev 001 10 13 SP Scientific 2013 ...
Страница 16: ...Getting Started FTS Systems FlexiCool 8 Rev 001 10 13 SP Scientific 2013 ...
Страница 22: ...Operation FTS Systems FlexiCool 14 Rev 001 10 13 SP Scientific 2013 ...
Страница 30: ...Maintenance FTS Systems FlexiCool 22 Rev 001 10 13 SP Scientific 2013 ...
Страница 37: ......