Rev 001, 10/13
19
© SP Scientific 2013
Chapter
5
Maintenance
ALWAYS REMOVE POWER FROM THE UNIT BEFORE PERFORMING MAINTENANCE
PROCEDURES.
NOTICE: ONLY QUALIFIED PERSONNEL SHOULD PERFORM MAINTENANCE ON THE
REFRIGERATION SYSTEM ITSELF, AND ONLY EPA CERTIFIED TECHNICIANS MAY
EVACUATE OR CHARGE REFRIGERANTS.
WARNING: HAZARDOUS VOLTAGES. RISK OF ELECTRIC SHOCK, DISCONNECT ALL POWER
BEFORE SERVICING THIS UNIT. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL.
HAZARDOUS MATERIAL/CONTACT HAZARD: THE CHEMICALS USED IN THIS UNIT ARE
CONTAINED WITHIN FULLY ENCLOSED SYSTEMS. NO MAINTENANCE ACTIVITIES
INVOLVING THE REFRIGERATION SYSTEM ARE TO BE PERFORMED BY THE END-USER.
HOT SURFACE: THE COMPRESSOR AND ASSOCIATED TUBING IN THE LOWER PORTION
OF THE UNIT ARE LIKELY TO BE HOT AND POSE A POTENTIAL BURN HAZARD.
Cleaning
The following procedure should be performed on a routine basis, from weekly to
monthly depending upon the operating environment.
Disconnect electric utilities.
1.
Wipe down all exterior cabinet surfaces with a damp cleaning cloth.
2.
Vacuum dust from the refrigeration condenser as well as all screens and
3.
openings.
Remove cabinet panels.
4.
Blow compressed air systematically from the top to bottom. Let air settle for
5.
several minutes and repeat.
Check for areas that have frosted during operation and insulate as required.
6.
Replace cabinet panels.
7.
Connect electric utilities.
8.
Содержание FTS FLEXICOOL IMMERSION COOLERS FC100
Страница 1: ...IMMERSION COOLERS FC50 FC100 OPERATOR S MANUAL FTS SYSTEMS FLEXICOOL IMMERSION COOLERS ...
Страница 2: ......
Страница 6: ...iv Rev 001 10 13 SP Scientific 2013 ...
Страница 16: ...Getting Started FTS Systems FlexiCool 8 Rev 001 10 13 SP Scientific 2013 ...
Страница 22: ...Operation FTS Systems FlexiCool 14 Rev 001 10 13 SP Scientific 2013 ...
Страница 30: ...Maintenance FTS Systems FlexiCool 22 Rev 001 10 13 SP Scientific 2013 ...
Страница 37: ......