ST2826 Operation manual
Ver1.0
◇
66
6.3.3Cable length factor
Cable length
0m
1m
D
1
(100kHz
max)1.5+0.015*f[kHz]
(100.1kHz
min)1.5+0.3*f[kHz]
6.3.4 Temperature factor
Temperature
Operation temperature=T
E
0.1*basic accuracy*(T-20)
Basic accuracy table 1(0.010V to 0.049V)
When range>=100
Basic accuracy
|
[ ]
[ ] |
(
)
[ ]
B
Zx
Range
A
Range
When range <100
Basic accuracy =
|
[ ] 10
[ ] |
(
)
[ ]
B
Range
Zx
A
Range
Zx:sample value of impedance Z
A,B:Basic accuracy factor
Accuracy factor of Z(%)
:
value A
,
B in group 1
Accuracy factor of
(
)
:
value A
,
B in group 2
When
1.001
f
MHz
,basic accuracy must multiply (f
[
MHz
]
+3)/4.
Table 6-5 Basic accuracy 1(0.010V to 0.049V)
*1
Range
20Hz-99.9Hz
100Hz-1kHz
1.001kHz-10kHz
10.01kHz-100kHz
100.1kHz-1MHz
1.001MHz-5MHz
1MΩ
100kΩ
A=1.5 B=1
A=1 B=0.5
A=1 B=0.5
A=2 B=0.3
A=0.5 B=0.1
A=0.5 B=0.1
10kΩ
A=0.8 B=0.5
A=0.5 B=0.2
A=0.5 B=0.2
A=1.5 B=0.5
A=0.5 B=0.01
A=0.2 B=0.02
A=0.2 B=0.02
A=0.6 B=0.05
3kΩ
A=0.8 B=0.5
A=0.5 B=0.2
A=0.5 B=0.2
A=1.5 B=0.5
A=0.5 B=0.01
A=0.2 B=0.02
A=0.2 B=0.02
A=0.6 B=0.05
Содержание ST2826
Страница 2: ......
Страница 10: ......
Страница 51: ...ST2826 Operation manual Ver1 0 41 Figure 3 15 ...
Страница 61: ...ST2826 Operation manual Ver1 0 51 ...
Страница 144: ...ST2826 Operation manual Ver1 0 134 Figure 9 8 the input circuit for control signal ...