ST2826 Operation manual
Ver1.0
◇
103
EXTernal
BUS
HOLD
:DELay <value>
MIN
MAX
:IMMediate
is used to trigger the instrument only one time.
command syntax
:
TRIGger[:IMMediate]
e.g.
:
WrtCmd(―TRIG‖);
:SOURce
is used to set the trigger source mode. Char ‗?‘ can be used to query the current
mode of the trigger source
command syntax
:
INTernal
TRIGger:SOURce EXTernal
BUS
HOLD
INTernal :triggered by the instrumetn itself automatically. It is the default setting.
EXTernal:triggered by the HANDLER interface.
BUS:triggered by RS232 or GPIB interface.
HOLD: triggered by the TRIGGER key in the front panel.
e.g.
:
WrtCmd(―TRIG:SOUR BUS‖);
query syntax
:
TRIGger:SOURce?
return
:
INT
EXT
<NL^END>
BUS
HOLD
:DELay
is used to set the delay time after the instrument is triggered. Char ‗?‘ can be used to
query the delay parameter.
command syntax
:
<value>
TRIGger:DELay
MIN
MAX
Содержание ST2826