ST2826 Operation manual
Ver1.0
◇
112
return
:
<NR3><NL^END>
:SPOT<n>:OPEN
is used to perform the open correction for the particular frequency
points(point 1,point 2 and point 3).
command syntax
:
CORRection:SPOT<n>:OPEN
<n>:
1:frequency point 1
2:frequency point 2
3:frequency point 3
e.g.
:
WrtCmd(―CORR:SPOT1:OPEN‖) ;perform the open correction for frequency
point 1.
:SPOT<n>:SHORt
is used to perform the short correction for the particular frequency
points(point1, point 2 and point 3).
command syntax
:
CORRection:SPOT<n>:SHORt
<n>:
1:frequency point 1
2:frequency point 2
3:frequency point 3
e.g.
:
WrtCmd(―CORR:SPOT1:SHOR‖) ;perform the short correction for point 1
:SPOT<n>:LOAD:STANdard
is used to perform the load correction for the particular
frequency points(point 1, point 2 and point 3) under the standard
reference value.
command syntax
:
CORRection:SPOT<n>:LOAD:STANdard <
REF.A
>,<
REF.B
>
<n>:
1:frequency point 1
2:frequency point 2
3:frequency point 3
<
REF.A
> : NR1
,
NR2 or NR3 format, is the standard reference value for the main
parameter
<
REF.B
> : NR1
,
NR2 or NR3 format, is the standard reference value for the
secondary parameter
e.g.:WrtCmd(―CORR:SPOT1:LAOD:STAN 100.7,0.0002‖)