ST2826 Operation manual
Ver1.0
◇
11
3.1.1Measure
In ST2826, two parameters of impedance component can be tested meanwhile
:
Primary
parameter and secondary parameter. The parameter is as below
:
Primary parameter
|Z|
(Module of impedance)
|Y| (Module of admittance)
L
(Inductance)
C
(Capacitance)
R
(Resistance)
G
(Conductance)
DCR (DC resistance)
Secondary parameter
D
(Dissipation factor)
Q
(Quality factor)
Rs (Equivalent series resistance)
Rp (Equivalent parallel resistance)
X
(Reactance)
B
(Admittance)
Ө
(Phase angle)
The test result of primary parameter and secondary parameter can be displayed with two lines
of large character. The primary parameter is displayed on the first line
,
and secondary
parameter is displayed on the second line.
Operation step of measure setting
:
1)
Move cursor to Function zone. The following softkeys are displayed
。
Cp-D
Cp-Q
Cp-G
Cp-Rp
More 1/7
2)
Select and press a softkey to set measure function. If the softkey you require is not
displayed on the page, press More 1/7 to display the following softkeys.
Cs-D
Cs-Q
Cs-Rs
More 2/7
3)
Select and press a softkey to set measure function. If the softkey you require is not
displayed on the page, press More 2/7 to display the following softkeys.