ST2826 Operation manual
Ver1.0
◇
98
FUNCtion:DEV<n>:MODE
PERCent
OFF
ABSolute:absolute deviation display
PERCent:percent deviation display
OFF : the direct display for the test value
<n>
:
char ‗1‘(49): set the deviation mode of the main parameters
char ‗2‘(50):set the deviation mode of the secondary parameters
e.g.
:
WrtCmd(―FUNC:DEV1:MODE ABS‖);
query syntax
:
FUNCtion:DEV<n>:MODE?
return
:
ABS
PERC
<NL^END>
OFF
:DEV<n>:REFerence
is used to set the standard deviation. Char ‗?‘ is used to query the
current standard deviation.
command syntax
:
FUNCtion:DEV<n>:REFerence<value>
<value>:NR1,NR2 or NR3 format
<n>
:
char ‗1‘(49):set the standard deviation of the main parameters
char ‗2‘(50):set the standard deviation of the secondary parameters
e.g.
:
WrtCmd(―FUNC:DEV1:REF 10‖);
query syntax
:
FUNCtion:DEV<n>:REFerence?
return
:
<NR3><NL^END>
:DEV<n>:REFerence:FILL
is used to set the standard deviation, which controls the
instrument only for one time. And copy the main and the secondary parameters as the standard
deviation.
command syntax
:
FUNCtion:DEV<n>:REFerence:FILL
<n>
:
char ‗1‘(49) or char ‗2‘(50) can be used to set the standard deviation for the
main parameters and the secondary parameters.
e.g.
:
WrtCmd(―FUNC:DEV1:REF:FILL‖);
8.1.10LIST subsystem command set
LIST
subsystem command set is used to set the List Sweep, the sweep point, the sweep mode