ST2826 Operation manual
Ver1.0
◇
109
MULTi
SINGle
SINGle:set or return the single channel mode.
MULTi: set or return the multi-channel mode.
e.g.
:
WrtCmd(―CORR:MEST MULT‖) ;set the instrument to multi-channel mode
query syntax
:
CORRection:MESTod?
return
:
SINGle
<NL^END>
MULTi
:OPEN
is used to perform the correction data of 41 preset test points.
command sytax
:
CORRection:OPEN
e.g.
:
WrtCmd(―CORR:OPEN‖)
:OPEN:STATe
is used to set the open correction function. Char ‗?‘ is used to query the state of
the open correction.
command syntax
:
ON
CORRection:OPEN:STATe OFF
1
0
char ‗1‘(49) :enable the open correction, which is equivalent to ON
char ‗0‘(48): forbid the open correction, which is equivalent to OFF.
e.g.
:
WrtCmd(―CORR:OPEN:STAT ON‖)
query syntax
:
CORRection:OPEN:STATe?
return
:
<NR1><NL^END>
:SHORt
is used to perform the short correction data of 41 preset test points.
command syntax
:
CORRection:SHORt
e.g.
:
WrtCmd(―CORR:SHOR‖)
:SHORt:STATe
is used to set the short correction function. Char ‗?‘ is used to query the state
of the short correction.
Command syntax
:
ON
CORRection:SHORt:STATe OFF
1