ST2826 Operation manual
Ver1.0
◇
61
(b)Press the softkey CORRECTION. CORRECTION page will be displayed.
(c)Move the cursor to OPEN field. ON, OFF and MEAS OPEN will be displayed in the
softkey area.
(d)Press ON to enable the open correction.
(e)Move the cursor to SHORT field. ON, OFF and MEAS SHORT will be displayed in the
softkey area.
(f)Press ON to enable the short correction.
(g)Move the cursor to LOAD field. ON and OFF will be displayed in the softkey area.
(h)Press ON to enable the load correction.
(i)Move the cursor to FUNC field. The current field will display Cp-D. Cp-D, Cp-Q, Cp-G,
Cp-Rp and MORE 1/6 will be displayed in the softkey area.
(j)Press Cp-D to select Cp-D.
(k)Move the cursor to FREQ 1 field. ON, OFF, MEAS OPEN, MEAS SHORT and MEAS
LOAD will be displayed in the softkey area.
(l)Press ON to enable the dot frequency correction for FREQ 1.
(m)Press[1][0][0], +100 will be displayed in the bottom of the screen, the usable units(Hz,
kHz and MHz) will be displayed. Press kHz. The value of FREQ1 will be changed to
100.000kHz which is the same to test frequency.
(n)Move the cursor to REF A field below FREQ1. Press [1][1], +11 will be displayed in the
bottom line of the screen. The usable units (p, n, μ, m and MORE 1/2) will be displayed in
the sotfkey area. Press n. the value of this field will be changed to 11.0000 nF .
(o)Move the cursor to B: field below FREQ 1. Press [0][.][0][0][5], +0.0005 will be displayed
in the bottom line of the screen, the usable units (p, n, μ, m and MORE 1/2) will be
displayed in the softkey area. Press *1, the value of this field will be changed to 0.00050.
(p)Move the cursor to FREQ 1 field. ON, OFF, MEAS OPEN, MEAS SHORT and MEAS
LOAD will be displayed in the softkey area.
(q)Keep the test fixture open, and keep your hands and other disturbance source far from the
test fixture. Press MEAS OPEN to perform open correction.
(r)Insert the shorting bar (ST26010) to the test fixture to ensure the shorting bar and the test
fixture contact close.
(s)Press the softkey MEAS SHORT to perform short correction.
(t)Insert the standard capacitance to the test fixture.
(u)Press the softkey MEAS LOAD to perform load correction.
Note:
(a) The displayed softkey information may be not identical to the one described in this book
because the software version is different.
(b)Load correction is effective to the instrument with the same specification. If the
specification is changed, the load correction should be done again.