ST2523 Series Operation Manual Ver1.2
45
This command switches the short function to ON or OFF.
Command syntax: :FUNC:SHORT ON
Query syntax: :FUNC:SHORT?
Return: 1(49) means ON.
0(48) meansOFF.
:ACFREQuency 50
60
This command is used to control the supply frequency.
Command syntax: :FUNC:ACFREQuency 50
Query syntax: :FUNC:ACFREQuency ?
Return:
50 the supply frequency is 50Hz
60 the supply frequency is 60Hz
7.1.3
APERture Subsystem Commands
The APERture subsystem commands are mainly used to set the measurement speed and the average times used
in measurement.
The APERture? query returns the current measurement speed and average times used in measurement.
Command Syntax:
FAST
APERture
MEDium
[,<value>]
SLOW
Where:
FAST: Fast: 100meas/sec
MEDium: Medium: 50meas/sec
SLOW : Slow: 6.25meas/sec
<value>
Average number ranging from 1 to 128(NR1).
For example: WrtCmd(“APER MED,55”);
Query syntax: APERture?
Return format:
FAST
MED
,<NR1><NL^END>
SLOW
7.1.4
TRIGger Subsystem Commands:
The TRIGger subsystem commands set the instrument trigger source and the delay time after triggering, and
triggers instrument measurement.