ST2523 Series Operation Manual Ver1.2
18
instrument finds the measurement result exceeding its range, and then ST2523 will stop testing. If the
“OFF” soft key is selected, the judge for the stop point will be switched off when scanning the trace.
3.7
<STATIS DISP>
Press [DISP] and then choose STATIS DISP to enter into the <STATIS DISP> page shown as below figure 3-8.
Figure 3-8 Statistic Display Page
The function of this page is to count measurement data, such as some engineering coefficients, the PASS/FAIL
rate. In this page, it is also workable to analyze the average value of results for multiple measurements. Details
are as below.
3.7.1
Mode
Move the cursor to it and select different modes through the soft key. There are two modes for comparison of
high-low value: ABS (resistance value, inductance value and voltage value of Hi/Lo) and % (percentage of
nominal value and Hi/Lo).
NOTE: the formula of percentage mode:
Upper limit value= nominal value*(1+ upper limit%)
Lower limit value= nominal value*(1+ lower limit%)
3.7.2
NOMA/NOMB
Move the cursor to it and input the data through the numerical keyboard.
A: nominal value of primary parameter
B: nominal value of secondary parameter
3.7.3
Number
Move the cursor to it and input the total measurement times of statistic analyses through the numerical keyboard.