other IEC/ISO Standards applicable to the
equipments.
2. Furthermore all configurations shall
comply with the system standard
IEC60601-1-1. Everybody who connects
additional equipment to the signal input
part or signal output part configures a
medical system, and is therefore,
responsible that the system complies with
the requirements of the system standard
IEC60601-1-1. If in doubt, consult the
qualified service personnel.
3. The leakage current could increase when
connected to other equipment.
4. For this particular equipment, all
accessory equipment connected as noted
above, must be connected to mains via an
additional isolation transformer
conforming with the construction
requirements of IEC60601-1 and
providing at least Basic Insulation.
5. This equipment generates, uses, and can
radiate radio frequency energy. If it is not
installed and used in accordance with the
instruction manual, it may cause
interference to other equipment. If this
unit causes interference (which can be
determined by unplugging the DC Cable
from the unit), try these measures:
Relocate the unit with respect to the
susceptible equipment. Plug this unit and
the susceptible equipment into different
branch circuit.
Consult your dealer. (According to standard
IEC60601-1-2 and CISPR11, Class B,
Group 1)