Guidance and manufacturer’s declaration - electromagnetic immunity
The UP-D711MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the UP-D711MD should assure that it is used in such as environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment-guidance
Portable and mobile RF
communications
equipment should be
used no closer to any part
of the UP-D711MD,
including cables, than the
recommended separation
distance calculated from
the equation appliance to
the frequency of the
transmitter.
Recommended
separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
d
= 1.2
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d
= 1.2
80 MHz to
800 MHz
d
= 2.3
800 MHz
to 2.5 GHz
Where
P
is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter manufacturer
and
d
is the
recommended separation
distance in meters (m).
Field strengths from
fixed RF transmitters, as
determined by an
electromagnetic site
survey,
a
should be less
than the compliance level
in each frequency
range.
b