GB
4
Important EMC notices for use in the medical environments
• The UP-D711MD needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the instructions for
use.
• The portable and mobile RF communications equipment such as cellular phones can
affect the UP-D711MD.
Warning
The use of accessories and cables other than those specified, with the exception of
replacement parts sold by Sony Corporation, may result in increased emissions or
decreased immunity of the UP-D711MD.
Warning
If the UP-D711MD should be used adjacent to or stacked with other equipment, it should be
observed to verify normal operation in the configuration in which it will be used.
Guidance and manufacturer’s declaration-electromagnetic emissions
The UP-D711MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the UP-D711MD should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic
environment-guidance
RF emissions
CISPR 11
Group 1
The UP-D711MD uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and
are not likely to cause any
interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The UP-D711MD is suitable
for use in all establishments,
including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable